Kit di analisi del contenuto di ioni Ferri nel siero
Nota: Prendi due o tre campioni diversi per la previsione prima del test.
Attrezzatura operativa: Spettrofotometro
Cat Number: BC1730
Misurare: 50T/48S
Componenti:
Reagente I: Polvere×2, storage at 4℃. Aggiungere 10 ml di acqua distillata prima dell'uso.
Reagent Ⅱ: Polvere×2, storage at 4℃. Aggiungere 313 μL glacial acetic acid and 10 ml di acqua distillata prima dell'uso. Soluzione standard: Liquido 3 ml×1, 1000 μmol/L Fe3+ standard solution, storage at 4℃. Add distilled water dilute 8 times to form a standard solution of 125 μmol/L before use.
Descrizione del prodotto:
Serum iron is the iron bound with transferrin in blood, which is often used to distinguish non-iron deficiency anemia and iron-deficiency anemia.
Fe3+ is reduced by sodium sulfite to Fe2+, which reacts with 2,2-dipyridine-bipyridine, have an absorption peak at 520 nm. According measure absorbance at 520 nm can reflect serum iron concentration.
Reagenti e attrezzature necessari ma non forniti.
Spettrofotometro, centrifuga, glacial acetic acid, adjusted transferpettor, 1 Cuvetta di vetro da ml, chloroform and distilled water.
Procedura:
- Preriscaldare lo spettrofotometro per 30 min, regolare la lunghezza d'onda su 520 nm, azzerare con il distillato
- Dilute standard solution to 125 μmol/L with distilled
- Aggiungere i reagenti con il seguente elenco:
Nome del reagente (μL) | Tubo vuoto (AB) | Provetta (A) | Tubo standard (COME) |
Acqua distillata | 400 | ||
Soluzione standard (125 μmol/L) | 400 | ||
Siero(plasma) | 400 | ||
Reagente I | 400 | 400 | 400 |
Reagente II | 400 | 400 | 400 |
Mescolare accuratamente, incubate in boiling water bath for 5 min, cooling liquid. Aggiungere 200 μL chloroform (required but not provided). Mescolare accuratamente, room temperature, 10000 rpm centrifuge for 10 min. Prendere 700 μL supernatant to 1 Cuvetta di vetro da ml. Measure absorbance at 520 nm. Recorded as AB, A, COME. |
Calcoli
Serum iron (μmol/L)=[Cs×(AT-AB)÷(As-AB)]= 125×(AT-AB)÷(As-AB)
Cs: Fe3+ standard solution, 125 μmol/L.
Nota:
- There is less iron in the serum, so the vessels (EP tubes) should be noted to avoid iron contamination.
- Reagent I and Reagent II are unstable. It needs to be prepared when the solution will be used, and the newly prepared reagent can only be used on the same day.
- When At- Ac>5, please dilute the Serum to appropriate concentration with distilled water.
Technical Specifications:
Minimum Detection Limit:3.212 µmol/ml
Linear Range:3.9-250 µmol/ml
Prodotti correlati:
BC2860/BC2865 Serum Total Iron Binding Capacity(TIBC) Kit di analisi
BC0720/BC0725 Blood Calcium Content Assay Kit